• YMC SEC columns and light scattering detection
    SEC in combination with MALLS can provide additional information: Chromatogram of BSA recorded with UV at 280 nm (black) and MALLS at 90° (blue) using a YMC-SEC MAB column.

Bioanalytical

YMC SEC columns and light scattering detection

Size exclusion chromatography (SEC) is not only used to separate compounds according to molar mass (MW) but also to gain information on the relative distribution of the MW of a sample.

SEC-LS: Method of choice

In combination with light scattering detection, SEC delivers absolute molar mass data for the characterisation of macromolecules such as proteins and antibodies. The greatest advantage is that there is no need for column calibration. Further, the signal intensity directly depends on the molar mass, injected mass and refractive index increment squared. This results in high signal intensities for higher MW materials, such as larger aggregates, which provides increased sensitivities which cannot be obtained by using a concentration detector (e.g. UV, RI). In contrast, light scattering (LS) is a technique that is suitable to generate absolute molar mass data for the characterisation of macromolecules such as proteins or antibodies.

Compatibility of SEC columns with light scattering

In order to combine SEC with MALLS, the SEC column has to be compatible with LS detection. Therefore, very high column purity is required to prevent spikes in the chromatogram caused by particles from the column itself. Such spikes cannot be seen in UV or RI detection.

YMC SEC column options for LS detection

YMC offers different SEC columns, with several pore and particle sizes. YMC-Pack Diol columns are available with 60 Å, 120 Å, 200 Å and 300 Å and with 3 μm and 5 μm for HPLC, while 2 μm columns for UHPLC proposes are available in combination with the 200 Å and 300 Å pore sizes. YMC-SEC MAB columns which are dedicated for monomer, fragment and aggregate analysis of antibodies are available with 250 Å and 3 μm particles.

All YMC SEC columns show a high compatibility with MALLS detection using typical eluents in BioLC such as phosphate buffers, as their noise levels are very low as required for such detection modes as shown and explained in detail in a Technical Note.


Digital Edition

Chromatography Today - Buyers' Guide 2022

October 2023

In This Edition Modern & Practical Applications - Accelerating ADC Development with Mass Spectrometry - Implementing High-Resolution Ion Mobility into Peptide Mapping Workflows Chromatogr...

View all digital editions

Events

SCM-11

Jan 20 2025 Amsterdam, Netherlands

Medlab Middle East

Feb 03 2025 Dubai, UAE

China Lab 2025

Feb 05 2025 Guangzhou, China

PITTCON 2025

Mar 01 2025 Boston, MA, USA

H2 Forum

Mar 04 2025 Berlin, Germany

View all events