Automated Analysis of Open-Access HPLC Instrumentation Metrics

Author: By S. Austin, D. Dunstan, J. Reilly, D. Wall, J. Ek, M. McKinnon on behalf of Novartis

Free to read

This article has been unlocked and is ready to read.

Download

All modern chromatographic software products gather information on each experiment and compile this data into text files. These files contain a wealth of information that can be used to track valuable metrics of instrument performance. By importing these metrics into data analysis software, researchers can easily visualise performance parameters such as retention times, peak heights, and intensities of samples across multiple systems simultaneously, while receiving updates in real time. This article discusses the utilisation of data analysis software for chromatographic system tracking in order to increase overall productivity and efficiency. 

Introduction

Analysis of QC standards for analytical instruments within a laboratory setting is a time consuming, albeit necessary process, often requiring dedicated technician support. As the number of dedicated open access instruments grows in an organisation, the routine task of maintaining and tracking instrument performance becomes resource intensive for analytical teams. Working in an open-access setting within Novartis Global Discovery Chemistry at NIBR Cambridge, there are over 20 analytical and preparative HPLC and LCMS systems that are set up to run automated QC standards multiple times a day to ensure optimal instrument performance. This workflow utilises analytical HPLC instruments to obtain an initial separation profile that aids chemists in method selection on a preparative scale. The retention times observed on analytical instruments are used to choose focused gradients for large scale preparative separations; therefore the accuracy and consistency of the analytical retention times is vital for the efficient purification of reaction products. 

In order to streamline the QC process, innovative ways to compile and monitor data generated within QC reports were examined. Longer term trends such as system utilisation or uptime are difficult to track and visualise, but these can play a large role in the efficiency of a chromatographic operation. Data analysis software has been utilised to visualise large data sets with many parameters in order to gain insight to discover and even predict future outcomes by analysing trends [1]. Data visualisation software from TIBCO (Spotfire) has been utilised within corporations ranging from scientific research institutes to financial institution data analytics departments, and was selected over various other programs for its proven versatility and visual data rendering capabilities [2]. 

 

Software utilised

Waters Masslynx data processing software v4.1 and TIBCO Spotfire v4.0.3.

 

QC Tracking

The main goal of monitoring our data was to track the chromatographic characteristics of standards run across multiple instruments. Instruments of the same type were standardised and QC datasets were collected over a period of several months. Averages of chromatographic data (retention time, peak height, peak width, absorbance, etc.) were recorded and used to determine a starting value within scatter plots. Limits were then determined to allow for unavoidable fluctuations in these values. These plots then give an easy way to visualise instrument performance and assign a QC pass or fail. 

By regularly checking these plots, it becomes much easier to monitor, and to determine when instruments require maintenance due to results that are not within specification. Previously undetected system changes were also evident through these visualisations. It was apparent on basic pH systems which utilise ammonium hydroxide as a modifier, that there is a clear trend in retention time shifts as the basic buffer additive appears to be evaporating over time.

Visualisations such as this have been beneficial in aiding the standardisation of open access analytical instruments, where the retention times have a critical impact on the determination of focused gradients for preparative purifications. When a retention time shifts on an analytical system due to system performance an incorrect purification methodology may be selected based upon this inaccurate analytical representation.

Although this is not a substitute to a researcher personally inspecting an instrument, the ability to collate performance parameters of instruments located within different laboratories or even different buildings can dramatically impact response times to malfunctioning HPLCs. 

Injection Tracking

The Waters Openlynx software allows chemists to input specified information fields upon logging in a purification sample.  Part of the login process has been configured to require the chemist to input project information prior to the injection of the sample. This aspect enables the ability to link project to metrics such as method and instrument selection, number of injections, and use per chemist, thus combining information that was not readily tracked previously. 

The “analytical scout-to-purification” workflow utilises focused gradients for preparative purifications based on retention times from analytical injections. Shown in Figure 3 is a plot of the varying focused gradient selections among two projects. Analysing metrics such as this can lead to the creation of customised methods within the open access framework for high demand specialty purifications. The spread of purification methods is very different based upon lipophilicity of samples. This can be seen in Figure 3 where for one project (green) the most utilised method was method 3 where compounds require low% acetonitrile gradients because they are more hydrophilic in nature. The other project (blue) utilised method 6 where compounds require more shallow high% acetonitrile gradients because they are more hydrophilic in nature.

 

Detector/ LC Performance

In monitoring the peak shape of standard runs, the performance of a specific instrument can be qualified. Peak height, width, and signal intensity are just a few parameters that can be tracked to ensure that detectors are functioning properly. These simple automated checks enable technicians to quickly focus on instruments where “hands on” maintenance is required on a daily basis. Instrument failures such as sample precipitation, column performance (peak width and retention time variation) and QC standard mass accuracy can be more simply tracked and instrument up-time can be more easily tracked. This gives a visual performance metric to all chemists and is shown in Figure 4 and demonstrates the usefulness of tracking performance metrics in this manner. 

Conclusion 

The initial goal of implementing data analysis software was for tracking QC runs, although a lasting impact was obtained from the unexpected additional metric information that this workflow gathered. The collation of specific project information and utilisation rates, allows for tailored chromatographic methods for high demand purifications. Daily operations become more streamlined as technicians are able to quickly identify and resolve issues remotely, while also performing simple preventative maintenance steps that minimises down time and increases column life. Predicting and planning for potential issues can be significantly improved by analysing column performance metrics through routine automated information tracking tests for chromatographic instruments.

References

(1) B. Evelson, The Forrester Wave™: Self Service Business Intelligence Platforms, Q2 2012. (Forrester Research Inc. 2012)

(2) T. Kucera, D.White, Predicitive Analytics for Sales and Marketing. (Aberdeen Group, A Harte-Hanks Company 2012)

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Chromatography Today - Buyers' Guide 2022

October 2023

In This Edition Modern & Practical Applications - Accelerating ADC Development with Mass Spectrometry - Implementing High-Resolution Ion Mobility into Peptide Mapping Workflows Chromatogr...

View all digital editions

Events

SCM-11

Jan 20 2025 Amsterdam, Netherlands

Medlab Middle East

Feb 03 2025 Dubai, UAE

China Lab 2025

Feb 05 2025 Guangzhou, China

PITTCON 2025

Mar 01 2025 Boston, MA, USA

H2 Forum

Mar 04 2025 Berlin, Germany

View all events